层析原理
|
|
|
|
|
|
|
|
|
|
## 使用1mM PB的目的
|
|
|
|
|
|
|
|
|
|
冲走OH离子,层析柱碱洗后使用注水冲洗,但NaOH冲柱会吸附OH根离子,用注水难冲走,所以使用1mM PB
|
|
|
|
|
|
|
|
|
|
## 脱盐的目的
|
|
|
|
|
|
|
|
|
|
原液中含有5mM PB,用1mM PB置换可以降低药液中的离子浓度,因为处方中辅料含有大量离子,可能会导致蛋白发生盐析,低盐药液会澄清
|
|
|
|
|
|
|
|
|
|
图谱异常
|
|
|
|
|
|
|
|
|
|
回顾历史
|
|
|
|
|
|
|
|
|
|
毛刺峰:
|
|
|
|
|
|
|
|
|
|
管路气泡
|
|
|
|
|
|
|
|
|
|
设备故障或不稳定(先确认硬件问题)
|
|
|
|
|
|
|
|
|
|
隔膜泵和紫外在一张桌子上,隔膜泵运行过程中震动影响紫外稳定性
|
|
|
|
|
|
|
|
|
|
新旧设备串联看差异
|
|
|
|
|
|
|
|
|
|
双峰:
|
|
|
|
|
|
|
|
|
|
柱子出现干柱(大气泡)
|
|
|
|
|
|
|
|
|
|
气泡来源
|
|
|
|
|
|
|
|
|
|
操作:
|
|
|
|
|
|
|
|
|
|
换管时操作或其他操作跑空引入
|
|
|
|
|
|
|
|
|
|
设备:
|
|
|
|
|
|
|
|
|
|
密闭性不好
|
|
|
|
|
|
|
|
|
|
物料:
|
|
|
|
|
|
|
|
|
|
水:含气量高
|
|
|
|
|
|
|
|
|
|
压力大、温度低水中含气量高
|
|
|
|
|
|
|
|
|
|
未调节水系统时,拆柱过程看到有气泡冒出,调节后看不出了
|
|
|
|
|
|
|
|
|
|
制水系统
|
|
|
|
|
|
|
|
|
|
提高制水设定限,减少注射用水储罐的空腔体积
|
|
|
|
|
|
|
|
|
|
降低循环泵压力,降低循环流速,降低压力
|
|
|
|
|
|
|
|
|
|
储水罐的空气滤芯透气量是否满足要求(加热干燥功能)
|
|
|
|
|
|
|
|
|
|
注射用水使用点不能放水放太快,出水温度在20~23℃之间,避免温度过低
|
|
|
|
|
|
|
|
|
|
怎么测定谁的含气量?使用生产前接水放置,观察产气量
|
|
|
|
|
|
|
|
|
|
原因:
|
|
|
|
|
|
|
|
|
|
平衡液:
|
|
|
|
|
|
|
|
|
|
原液:
|
|
|
|
|
|
|
|
|
|
20年原液没有2~8℃冻存,21年原液要冻存后上样
|
|
|
|
|
|
|
|
|
|
操作:除NaOH不能过滤芯外,全部过滤芯
|
|
|
|
|
|
|
|
|
|
展开
|
|
|
|
|
|
|
|
|
|
水溶液制剂也要主要水的含气量→配制→分装到各个小瓶中
|
|
|
|
|
|
|
|
|
|
# 措施
|
|
|
|
|
|
|
|
|
|
原则:G25跟制剂投料的PB分开,配方体现按原来的不变
|
|
|
|
|
|
|
|
|
|
制剂配10mM投料,纯化配0.1M母液配制30L左右够1个月使用的量,有效期要做到1个月
|
|
|
|
|
|
|
|
|
|
实际多投的16g另外买1瓶回来,每次称20g补过来,第一次的可以做损耗
|
|
|
|
|
|
|
|
|
|
变更与纳音:提高浓度,减少配制次数,提高生产效率
|
|
|
|
|
|
|
|
|
|
统计相关涉及文件和工作
|